S129 Methane Analyzer 500ppm

Home Products Gas Analysis and Environment S129 Methane Analyzer 500ppm
S129 Methane Analyzer 500ppm

Qubit S129 Methane Analyzer is a non-dispersive infrared (NDIR) gas analyzer that measures methane levels in a flowing gas at 0-500ppm range.  The instrument is equipped with a digital display and an analog output signal (0 – 5 V) for data logging data by a computer with our optional software and data interface.

The S129 Methane Analyzer is a flow through gas analyzer therefore gas should be supplied under pressure at flow rates not to exceed 650ml/min.  An optional built in gas pump may be provided ( S129-P) for easy delivery of the gas to the sensor.  Additionally, a mass flow monitor may also be included for measurements of flow rate through the sample chamber (S129-PF).  The S129 Methane analyzer is housed in a weather proof case with temperature control for enhanced stability of the signal and is ideally suited for both lab and field research (with an optional battery pack).

  • Flow-through infra red gas sensor
  • 0-500ppm standard range
  • Optional enhanced range available (0-100ppm)
  • Linear response
  • Easy two point calibration
  • Digital display
  • 0-5V analog output
  • Optional built in gas pump (S129-P)
  • Optional build in gas pump and flow monitor (S129-PF)
  • Temperature controlled case for enhanced signal stability
  • Optional battery pack available for field operation
  • Bioreactors and Biofuels research
  • Animal gas exchange
  • Livestock emissions
  • Land emissions
  • Operating principle Non-dispersive infrared
  • Gas sampling mode Flowing gas stream, sealed chamber
  • Maximum gas flow rate 650 mL/min
  • Measurement range (LCD display) 0 – 500ppm
  • Optional enhanced range 0-100ppm
  • Accuracy: 0.5% FS error
  • Repeatability (at stable atm press and temp) ± 2ppm CH4
  • Response time (@ 250 mL/min; to 95% of final value)   ca. 25 sec
  • Warm up time (@ 22C)  1hr
  • Output (linear) for  1 – 4 VDC for 0 – 500ppm
  • Calibration adjustments:  Zero and Span
  • Two point calibration
  • Operating ambient temperature range  0 to 35C
  • Built in heater with temperature controller
  • Storage temperature range   -40 to 60C
  • Operating pressure range ±1.5% local mean pressure
  • Humidity range 5 to 95% RH, non-condensing (essential to dry the gas stream)
  • Pressure dependence +0.19% reading per mm Hg
  • Power requirements  12 VDC via 120/240 VAC, 60/50 Hz adapter
  • Current requirements  125 mA average, 450 mA peak
  • Dimensions (cm) (H x W x D: 40 x 31 x 18)
  • Weight 4.2 kg
  • Warranty  1 year limited

Contact Qubit Systems